Working
Environment |
|
Temperature:10℃~35℃
Relative Humidity:
≤80﹪
Atmospheric Pressure:86kPa
~106kPa |
|
Transport and Storage Environment |
Temperature:-10℃~40℃ Relative Humidity:≤80﹪
Atmospheric Pressure:50kPa~106kPa |
|
Power Supply Requirements |
Voltage: (single-phase) 220V±22V;20A
Frequency:50Hz±1Hz |
|
Display Function |
Display content:ECG waveform, Finger Pleth waveform,
inflation/deflation signal, SPO2,
pressure value, heart rate, NIBP, D/S ratio and
residual treatment time.
Scanning speed:25mm/s or 50mm/s. |
|
Operation Function |
|
Patient Protection |
◆The
System has following Functions
◆which
are ECG Amplitude
◆Finger
Pleth Amplitude
◆Inflation/Deflation
starting time Adjustment
◆Air
Pressure Adjustment
◆Treatment
Time Adjustment
◆Solenoid
Valve control
◆Compressor
Control
◆SPO2
Alarm, Blood Pressure Alarm
◆Patient
Information, and Printing. |
|
◆Automatic
inflation stop and pressure release on early
extra systole.
◆
Automatic pressure release whenever an extra
systole or other indicator of an
arrhythmia or aberrant activity is detected.
◆Automatic
treatment stop and pressure release if power
outage occurs.
◆Automatic
pressure release if air reservoir tank
pressure is more than 59kPa.
◆Automatic
high-pressure limit. Automatic release if
the pressure is 5 kPa over
setting.
◆Automatic
stop and pressure release if heart rate is
goes out of range 35~125bpm.
◆Automatic
pressure release if ECG electrodes fell down
during treatment.
◆Emergency
stop button. |
|
|
|
ECG
Amplifier |
Input
impedance: ≥4MΩ Maximum Augment
Gain:≥60dB Common mode rejection ratio(CMRR):
≥90dB Minimum ECG detection voltage:
≤0.25mV |
|
|
Pulse Detection |
|
Method:use Finger Plethysmograph Maximum
Augment:≥60dB(8 grades adjustable) |
|
NIBP
Measure |
NIBP required pressure range |
|
60~250mmHg(default
is 160 mmHg) |
NIBP measuring interval |
|
5~50mins(default
is 20mins) |
NIBP( high pressure) warning range |
|
upper limit:160~210mmHg(default is 200 mmHg) |
|
lower limit:0~120mmHg(default is 100 mmHg) |
NIBP (high pressure) Stop Treatment range |
|
Upper limit:210~230mmHg(default is 230 mmHg) |
|
Lower limit: 70~90mmHg(default is 90 mmHg) |
|
|
|
SPO2
Oxygen Saturation |
◆Oxygen
saturation range: 0 ~ 100﹪。
◆Oxygen
saturation warning range:50﹪ ~ 99﹪。
◆Accuracy
SpO2:± 1 digits for using the Finger Clip Sensor.
◆Oxygen
saturation is lower than warning value, the System
will alarm.
◆Oxygen
saturation is lower than 90%, record it into file.
◆Oxygen
saturation is lower than 80%, stop the compressor
automatically.
|
|
|
Triggering Method |
◆Internal
trigger:the System produce simulate square wave,
range from 10~170bpm
◆External
trigger:R-wave of patient ECG
◆External
trigger ratio: 1:1 or 1:2
◆ECG
trigger range: 35~125 bpm ±1 bpm |
|
|
Filter Technique |
Low-pass digital filter assorted with double T
filter eliminates 50/60Hz and high-frequency
interference, as well as 40Hz body muscle
electricity interference. |
|
|
Treatment Pressure Range |
|
Adjustable from :0~50kPa(classified in
20grades;2.5kPa /stage) |
|
Cuffs Inflation Method |
Three
classes sequence inflation from calves, low thighs
and upper thighs. |
|
Table Safe Working Load |
400
lb (180 kg) |
|
Treatment Timer |
Treatment Time setting range is 0~120 mins, the
default is 60 mins. |
|
System Protection |
The
circuit breaker: main fuse is
15A、computer system fuse is 3A. |
|
|
Continue Working Time |
More
than 8 hours per day in the regulate situation. |
|
|
Equipment Dimensions and Weight |
Length |
Width |
Height |
Net Weight |
2170mm |
930mm |
650mm |
265kg |
|